or
Contact Us
BOOK A CALL
Book Call On Your Favorite Time
Code
🗘

By Signing Up. I agree to TradeKey.com Terms of Use, Privacy Policy, IPR and receive emails related to our services

UVTC2 UV Test Chamber for Two PV Modules

  • Pakistan Pakistan  
  • Quantity Required: 1 Unit
  • Annual Purchase Volume: 1 Unit
  • Last Updated:
    Posted on: 07 Nov 2024

  • Buying Frequency: One time
  • Payment Terms: T/T
  • Destination Port: karachi pakistan

RFQ Details

UV chamber is required to determine the resistance of photovoltaic modules against
UV-Radiation, according to the standards IEC 61215:2016 and IEC 61215-2:2021. The
design of the radiation concept addresses the specific requirements of the test
Section VII " Schedule of Requirements 89

procedures as explained in paragraph MQT 10 „UV Preconditioning Test“ and MST54
“UV Test”. For the generation of the UV light we use special UV tubes which generate a
solar spectrum a spectrum in accordance with IEC 61215. Technical specifications of
the UV radiation system are as follows:
Inhomogeneity of irradiance: Better 15%
Radiation Intensity: The radiation Intensity within a wavelength range of 280 nm up to
400 nm does not exceed 250 W/m² (maximum 5-times of the natural solar UV
intensity)
Spectral Distribution: No significant Intensities < 280 nm
Radiation (Dosage): During the Tests the PV module shall receive the following
radiation dosage; 15 kWh/m² within 280-400 nm; minimum 3-10% within 280-320 nm
(UVB)
Available Test Areas: Two Modules: 2 x 2600 x 1400 mm

Following are the accessories/ parts of the equipment
UVTC2 - UV test chamber for two PV modules: The chamber can be used for
accelerated UV aging tests and is specially designed for MQT 10: UV preconditioning
test"" according to IEC 61215-2:2021.
UVTC2 - Monitoring equipment for continuous monitoring of intensity and module
temperature

UVTC2 - Set of spare lamps (for one module test area)
UVTC2 Option - Calibration kit
UVTC2 Option - Calibration package ISO17025
UVTC2 Option - Feedthrough
UVTC2 - Additional temperature sensors with data acquisition unit to monitor the
module temperature.
Setup must be supplied with all necessary accessories required to run the whole
setup.


Refine by popular tags